4.19.10 :DIGits <n>
[:SENSe[1]]:CURRent:AC:DIGits <n> Specify resolution for ACI
[:SENSe[1]]:CURRent[:DC]:DIGits <n> Specify resolution for DCI
[:SENSe[1]]:VOLTage:AC:DIGits <n> Specify resolution for ACV
[:SENSe[1]]:VOLTage[:DC]:DIGits <n> Specify resolution for DCV
[:SENSe[1]]:RESistance:DIGits <n> Specify resolution for 2
[:SENSe[1]]:FRESistance:DIGits <n> Specify resolution for 4
[:SENSe[1]]:FREQuency:DIGits <n> Specify resolution for FREQ
[:SENSe[1]]:TEMPerature:DIGits <n> Specify resolution for TEMP
Parameters <n> = 4 3½ digits for DCI, ACI, ACV, DCV, 2, 4
3½ digits (1° resolution) for TEMP
4 digits for FREQ
= 5 4½ digits for DCI, ACI, ACV, DCV, 2, 4
4½ digits (0.1° resolution) for TEMP
5 digits for FREQ
= 6 5½ digits for DCI, ACI, ACV, DCV, 2, 4
5½ digits (0.01° resolution) for TEMP
= 7 6½ digits for DCI, ACI, ACV, DCV, 2, 4
6½ digits (0.001° resolution) for TEMP
= 8 7½ digits for DCI, ACI, DCV, ACV, 2, 4, TEMP
= 9 8½ digits for DCI, ACI, DCV, ACV, 2, 4
= DEFault 7½ digits for TEMP
6½ digits for DCI, ACI, DCV, ACV, 2, 4
5 digits for FREQ
= MINimum 3½ digits for DCI, ACI, ACV, DCV, 2, 4
3½ digits (1° resolution) for TEMP
4 digits for FREQ
= MAXimum 8½ digits for DCI, ACI, DCV, ACV, 2, 4
7½ digits for TEMP
5 digits for FREQ
Formats :curr:ac:dig <n> :volt:ac:dig <n>
:curr[:dc]:dig <n> :volt[:dc]:dig <n>
:res:dig <n> :freq:dig <n>
:fres:dig <n> :temp:dig <n>
Defaults Power-up Saved power-on setup
*RST 7½ digits for TEMP
6½ digits for DCI, ACI, DCV, ACV, 2, 4
5 digits for FREQ
:SYSTem:PRESet Same as *RST

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz