EasyManuals Logo
Home>Keithley>Multimeter>2001

Keithley 2001 User Manual

Keithley 2001
488 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
Front Panel Operation
3-3
Table 3-1
Data checked on power-up
Data Type of storage Memory option
IEEE-488 address
Power-on default
Electrically-erasable PROM
Electrically-erasable PROM
STD, MEM1, MEM2
STD, MEM1, MEM2
Calibration constants
Calibration dates
Electrically-erasable PROM
Electrically-erasable PROM
STD, MEM1, MEM2
STD, MEM1, MEM2
Instrument setups 1 in electrically-erasable PROM
4 more in non-volatile RAM
9 more in non-volatile RAM
STD, MEM1, MEM2
MEM1
MEM2
Reading buffer (Volatile RAM)
Non-volatile RAM
STD
MEM1, MEM2
Note: STD is standard memory, MEM1 is memory option #1, MEM2 is memory option #2.
Table 3-2
Power-up error messages
Message Action
Error +515,
Calibration dates lost
The cal dates are set to factory default values, but they are not stored into EEPROM. To do
this, perform a comprehensive calibration.
Error +514,
DC calibration data lost
DC cal constants are set to factory default values, but they are not stored into EEPROM. To
do this, perform a comprehensive calibration.
Error +513,
AC calibration data lost
AC cal constants are set to factory default values, but they are not stored into EEPROM. To
do this, perform a comprehensive calibration.
Error +512,
Power-on state lost
Power-on defaults are reset to factory defaults (bench) and stored into EEPROM.
Error +511,
GPIB address lost
GPIB address is reset to factory default (16) and stored into EEPROM.
Error +510,
Reading buffer data lost
The reading buffer controls are reset to factory defaults, but they are not stored into
NVRAM. To do this, store readings in the buffer.
Error -314,
Save/recall memory lost
Instrument setup is reset to bench defaults and stored into EEPROM.
Notes:
1. Any of these error conditions may occur the first time a unit is turned on or after replacing the firmware.
2. Error +510, Reading buffer data lost, applies to units with optional memory.
3. Error +512, Power-on state lost, may occur the first time the unit is powered-up after replacing the MEM2 memory option with MEM1.

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
BrandKeithley
Model2001
CategoryMultimeter
LanguageEnglish

Related product manuals