IEEE-488 Reference
4-178
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Questionable Event
0 = Disable (Mask) Questionable Event
Events : Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
Cal
B13 - B9 B8B15 B14
(2 )
8
256
(2 )
14
16384
0/10/1
Warn
B7 - B5
0
Temp
B4
(2 )
4
16
0/1
B3 - B0
F
igure 4-29
Questionable Event Enable Register
F
igure 4-30
Operation Event Enable Register
Idle Arm Trig
B15 - B11 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1 B0
(2 )
10
1024
(2 )
9
(2 )
6
(2 )
5
(2 )
1
(2 )
0
64 32
0/1 0/1 0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Operation Event
0 = Disable (Mask) Operation Event
Set
2
0/1
Events : Idle = Idle state of the 2001
Calc = Calculating
Arm = Waiting for Arm
Trig = Waiting for Trigger
Set = Settling
Cal = Calibrating
Calc Cal
512 1
0/1 0/1

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz