IEEE-488 Reference
4-83
Query :ENABle? Query state of display
Short-form format: :disp:enab?
Response message: 1 (on) or 0 (off)
Description This command is used to enable and disable the front panel display circuitry. Disabling the dis-
play circuitry allows the instrument to operate at a higher speed. While disabled, the display will
be frozen with the following message:
FRONT PANEL DISABLED
Press LOCAL to resume.
As reported by the message, all front panel controls (except LOCAL) will be disabled. Normal
display operation can be resumed by using the :ENABle command to enable the display or by
pressing the LOCAL key.
Programming example 10 OUTPUT 716; “:disp:enab off; enab?”
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10 Disables the display, and then queries the state of the display.
Line 20 Addresses the Model 2001 to talk.
Line 30 Displays (on the CRT) the programmed state of the display (0; off).

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz