IEEE-488 Reference
4-109
[:SENSe[1]]
:CURRent[:DC]
:DIGits?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe] <b>
[:STATe]?
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:METHod <name>
:METHod?
Query resolution.
Path to configure and control the filter.
Select filter type: MOVing, REPeat.
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Path to configure and control advanced filter.
Specify noise tolerance level (0 to 100 percent).
Query noise tolerance level.
Enable (1 or ON) or disable (0 or OFF) advanced filter.
Query state of advanced filter.
Enable (1 or ON) or disable (0 or OFF) filter.
Query state of digital filter.
Enable (1 or ON) or disable (0 or OFF) auto filter.
Enable and then disable auto filter.
Query auto filter.
Select measurement technique: NORMal, ICIRcuit.
Query method.
4.19.11
4.19.15
:VOLTage:AC
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
:COUPling AC|DC
:COUPling?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO ONCE
:ULIMit <n>
:ULIMit?
:LLIMit <n>
:LLIMit?
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
Path to configure AC voltage.
Specify integration rate in seconds (166.67e-6 to 200e-3).
Enable (1 or ON) or disable (0 or OFF) auto aperture.
Enable and then disable auto aperture.
Query auto aperture (0 or 1).
Query aperture (integration rate).
Specify integration rate (number of line cycles; 0.01 to 10).
Enable (1 or ON) or disable (0 or OFF) auto NPLC.
Enable and then disable auto NPLC.
Query auto line cycle integration (0 or 1).
Query line cycle integration rate.
Specify input coupling.
Query input coupling.
Path to configure measurement range:
Select range (0 to +775).
Query range.
Enable (1 or ON) or disable (0 or OFF) auto range.
Set range based on present input signal.
Specify upper limit for auto range (0 to +775).
Query upper limit.
Specify lower limit for auto range (0 to +775).
Query lower limit.
Query auto range.
Specify reference (REL) value (-1100 to +1100).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Use input signal as reference.
Query reference (REL) value.
4.19.5
4.19.6
4.19.7
4.19.8
4.19.9
Table 4-12 (cont.)
SENSe command summary
Command Description Paragraph