EasyManuals Logo

Keithley 2001 User Manual

Keithley 2001
488 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #332 background imageLoading...
Page #332 background image
IEEE-488 Reference
4-168
Description
These commands are used to set the logic levels for the digital output port. The actual level (high
or low) of a digital output line depends on its programmed polarity (see Output subsystem). If
the polarity of a line is active-high, then that line will be high when it is set to be true (1 or ON),
and low when it is set to be false (0 or OFF). Conversely, if polarity is active-low, the line will
be low when it is set to be true, and high when it is set to be false.
Programming example
10 OUTPUT 716; “:sour:ttl on; ttl?”
20 ENTER 716; A$
30 PRINT A$
40 END
Line 10 Sets output line #1 to true, and then queries the state of the output line.
Line 20 Addresses the Model 2001 to talk.
Line 30 Displays the state of output line #1 (1; on).

Table of Contents

Other manuals for Keithley 2001

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2001 and is the answer not in the manual?

Keithley 2001 Specifications

General IconGeneral
Model2001
CategoryMultimeter
ManufacturerKeithley
TypeDigital Multimeter
Digits of Resolution7.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
ConnectivityGPIB, RS-232
DisplayVFD
Resistance Accuracy0.008% + 2.0 mΩ
Frequency Accuracy±0.01%
DC Current Range10 μA to 3 A
AC Current Range10 μA to 3 A
Power Supply100 V to 240 V AC, 45 Hz to 66 Hz