SCPI Conformance Information
J-14
[:SENSe[1]]
:FRESistance
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:ADVanced
:NTOLerance <n>
:NTOLerance?
[:STATe] <b>
[:STATe]?
[:STATe] <b>
[:STATe]?
:AUTO <b>
:AUTO ONCE
:AUTO?
:FREQuency
:COUPling AC|DC
:COUPling?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:DIGits <n>
:DIGits?
:SOURce <name>
:SOURce?
:THReshold
:CURRent
:RANGe <n>
:RANGe?
:LEVel <n>
:LEVel?
:VOLTage
:RANGe <n>
:RANGe?
:LEVel <n>
:LEVel?
:TTL
:TEMPerature
:APERture <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:APERture?
:NPLCycles <n>
:AUTO <b>
:AUTO ONCE
:AUTO?
:NPLCycles?
Path to configure and control the filter.
Select filter type: MOVing, REPeat.
Query filter type.
Specify filter count (1 to 100).
Query filter count.
Path to configure and control advanced filter.
Specify noise tolerance level (0 to 100 percent).
Query noise tolerance level.
Enable (1 or ON) or disable (0 or OFF) advanced filter.
Query state of advanced filter.
Enable (1 or ON) or disable (0 or OFF) filter.
Query state of digital filter.
Enable (1 or ON) or disable (0 or OFF) auto filter.
Enable and then disable auto filter.
Query auto filter.
Path to configure frequency:
Specify input coupling.
Query input coupling.
Specify reference (REL) value (0 to 15e6).
Enable (1 or ON) or disable (0 or OFF) REL.
Query state of REL (0 or 1).
Use input signal as reference.
Query reference (REL) value.
Specify measurement resolution (4 or 5).
Query resolution.
Select source: CURRent, VOLTage.
Query source.
Path to configure threshold:
Path to set current threshold:
Specify range (0 to 1).
Query range.
Specify threshold level.
Query threshold level.
Path to set voltage threshold:
Specify range (0 to 1000).
Query range.
Specify threshold level.
Query threshold level.
Set threshold TTL level.
Path to configure four-wire resistance:
Specify integration rate in seconds (166.67e-6 to 200e-3).
Enable (1 or ON) or disable (0 or OFF) auto aperture.
Enable and then disable auto aperture.
Query auto aperture (0 or 1).
Query aperture (integration rate).
Specify integration rate in number of line cycles (0.01 to 10).
Enable (1 or ON) or disable (0 or OFF) auto NPLC.
Enable and then disable auto NPLC.
Query auto line cycle integration (0 or 1).
Query line cycle integration rate.
Table J-2 (cont.)
Syntax of non-SCPI commands implemented by Model 2001
Command Description