EasyManuals Logo

Keysight Technologies B1505A Configuration Guide

Keysight Technologies B1505A
298 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
4-8 Keysight B1505A Configuration and Connection Guide
N1272A and N1273A Connection Guide
Output Connection
Output Connection
The following parts are available for mounting a device under test (DUT) on the test
fixture. Prepare the part suitable for your DUT and attach it to the test fixture. For details,
see the following sections.
"3-pin Inline Package Socket Module" on page 4-9
"N1273A-011 Universal Socket Module" on page 4-11
"N1273A-013 Curve Tracer Test Adapter Socket Module" on page 4-14
To Connect DUT
WARNING Set the instrument output off before connecting or disconnecting connection wire.
Press the B1505A front panel
Stop key to set the output off. And confirm that the
B1505A front panel High Voltage indicator is not lit.
WARNING Do not connect or disconnect your DUT while the B1505A is applying voltage or
current. Otherwise, the DUT may be damaged.
Do not touch the DUT soon after measurement. Have enough time for discharge and
radiation.
CAUTION Do not touch the terminals of the cable and wire. Oil, perspiration, and dirt prevent good
electrical contact, deteriorate insulation, and degrade measurement accuracy.
CAUTION Never connect the HVSMU Force and Guard terminals to any output, including circuit
common, chassis ground, or any other module terminal. Connecting other module may
damage the connected module.
1. Set the instrument output off.
2. Open the fixture cover.
3. Set a socket module on the test fixture.
For using the 3-pin inline package socket module, also see "3-pin Inline Package
Socket Module" on page 4-9.
For using the universal socket module, also see "N1273A-011 Universal Socket
Module" on page 4-11.
For using the curve tracer test adapter socket module, also see "N1273A-013 Curve
Tracer Test Adapter Socket Module" on page 4-14.
4. Set the DUT on the socket.
5. Close the fixture cover.
To apply more than ± 42 V, close the fixture cover. Otherwise, the interlock function
will stop the B1505A output over ± 42 V.

Table of Contents

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals