5-14 Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect High Voltage R-Box
To Connect High Voltage R-Box
Keysight N1262A resistor box is used to reduce the risk of device breakdown or prevent
SMU from oscillation. The N1262A should be installed in your DUT interface as shown in
Table 5-1 on page 5-5.
N1262A-001: 1 M (± 3000 Vdc) resistor, HV(jack)-SHV(jack)
N1262A-002: 100 k (± 3000 Vdc) resistor, HV(jack)-SHV(jack)
N1262A-010: 1 k (± 200 Vdc) resistor, Triaxial(f)-Triaxial(f)
N1262A-011: 1 k (± 3000 Vdc) resistor, HV(jack)-SHV(jack)
N1262A-020: Universal R-box, Triaxial(f)-Triaxial(f)
N1262A-021: Universal R-box, HV(jack)-SHV(jack)
N1262A-023: Universal R-box, UHV(jack)-UHV(jack)
N1262A-036: 50 termination adapter
Universal R-box is kind of a do-it-yourself kit for installing a resistor you want. To install
the resistor, see "To Use Universal R-Box" on page 5-15.
N1262A-036 termination adapter is a BNC(f)-BNC(m) adapter which contains 50
resistor between signal line and shield. If a SMU is used as the gate drive for the
measurement using HVMCU, UHVU, or UHCU, the SMU may cause oscillation. Then
this adapter should be connected between the gate drive output and a DUT terminal. This
will be effective for preventing oscillation.
Table 5-8 To Connect N1262A Input
Connect from Required cables N1262A terminals
HVSMU
Force 16493T-001 HVSMU cable, 1.5 m From HVSMU
N1261A-001
Force 16494A-001 Triaxial cable, 1.5 m, for N1262A-010 or
N1262A-020
From SMU Force
N1261A-003
Force 16493T-001 HVSMU cable, 1.5 m
UHVU
High UHV cable of 16493V-001 UHVU cable, 1.5 m From UHV
N1262A
Opt 001
High Voltage Series R 1M
From HVSMU
Signal
Circuit
Common
3kV Max
䞉䞉䞉䞉
Output
!
!
3kV Max
䞉䞉䞉䞉
Agilent
!
!
Max : 9W
N1262A
Opt 001
High Voltage Series R 1M
From HVSMU
Signal
Circuit
Common
Signal
Circuit
Common
3kV Max
䞉䞉䞉䞉
3kV Max
䞉䞉䞉䞉䞉䞉䞉䞉
Output
!
!
3kV Max
䞉䞉䞉䞉
!!
!
3kV Max
䞉䞉䞉䞉
3kV Max
䞉䞉䞉䞉䞉䞉䞉䞉
Agilent
!
!
!!
!
Max : 9W

Table of Contents

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals