5-46 Keysight B1505A Configuration and Connection Guide
Connection Guide for Wafer Prober and Your Own Test Fixture
To Connect Gate Charge Adapter
Figure 5-23 Typical Connection of N1275A Gate Charge Measurement Adapter
G
S
HV SMU
GN DU
UHC Input
VControl
IControl
Fo r Gat e
For N1265A
For N1265A
For Drain
HF
LF
HS
LS
G
ate
N1265A
Gate
Selector
High
Low
N1254A-524
D
Ga te F
Ga te S
B1512-61741
(Included in the N1275A)
N1275-61701
(Included in the N1275A)
Interlock
Shielding Box
Shielding Box
HVSMU
HC/MCSMU
HC/MCSMU
MCSMU
GNDU
F
S
F
S
F
S
B1505A
16493T (Included in the HVSMU)
16494A (for MCSMU, included in the MCSMU)
16493L (Included in the B1505A)
16493G-001 Digital I/O cable (Included in the N1265A)
16493J (Included in the B1505A)
HP/MPSMU
F
S
MCSMU
F
S
N1275A
HF From N1254A
To DUT
LS From N1254A
To DUT
F
S
MCSMU
16494A (Included in the MCSMU)
N1261A-003
Output F
S
F
S
HPSMU
16494A (Included in the MPSMU)
Connecon for High Current gate
charge measurement
G
S
HF
LF
HS
LS
G
ate
A
N1254A-524
D
B1512-61741
(Included in the N1275A)
N1275A
HF From N1254A
To DUT
LS From N1254A
To DUT
F
S
MCSMU
Connecon for High Voltage gate
charge measurement

Table of Contents

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals