EasyManua.ls Logo

Keysight Technologies N9030B

Keysight Technologies N9030B
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
442 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area (Options 503, 508, 513, 526)
Replacement
1. Refer to Figure 16-8. Position the attenuator in the bracket so that the
ribbon connector end is “up”.
2. Replace the two screws (0515-0372) that attach the attenuator to the
bracket. Torque to 9 inch-pounds.
3. Position the attenuator and bracket into the chassis and attach with the
two screws (0515-0372). Torque to 9 inch-pounds.
4. Replace the ribbon cable and semi-rigid cables W2, W3, W4, or W6 to the
attenuator. Torque the semi-rigid cables to 10 inch-pounds.
5. Refer to Figure 16-7. Position the RF bracket onto the chassis and replace
the sixteen screws (0515-0372). Torque to 9 inch-pounds.
6. Replace the front panel. Refer to the Front Frame Assembly replacement
procedure.
7. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.

Table of Contents

Other manuals for Keysight Technologies N9030B

Related product manuals