N9030B PXA Signal Analyzer Service Guide 481
Assembly Replacement Procedures
RF Area (Options 544, 550)
Replacement
1. Refer to Figure 16-47. Place the switch onto the bracket and replace the
five screws (3) (0515-0372). Torque to 9 inch-pounds.
2. Place the switch/bracket into place into the chassis and replace the two
screws (0515-0372). Torque to 9 inch-pounds starting with the screw
closest to the front of the instrument.
3. Refer to Figure 16-45. Replace the semi-rigid cables W6, W8, and W56.
Torque to 10 inch-pounds.
4. Replace the ribbon cable W7. Ensure locking tabs on the sides of the
connector are engaged.
5. Refer to Figure 16-43. Position the RF bracket onto the chassis and
replace the sixteen screws (0515-0372). Torque to 9 inch-pounds.
6. Replace the front panel. Refer to the Front Frame Assembly replacement
procedure.
7. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.