EasyManua.ls Logo

Keysight Technologies N9030B

Keysight Technologies N9030B
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
486 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area (Options 544, 550)
Figure 16-52 YTO Screws
Replacement
1. Refer to Figure 16-52. Place the YTO into the chassis. Replace the two
screws (1) (0515-0372). Torque to 9 inch-pounds.
2. Figure 16-51. Replace the cables W20. Torque to 10 inch-pounds.
3. Replace the ribbon cable W23. Ensure the locking latch is engaged.
4. Refer to Figure 16-43. Position the RF bracket onto the chassis and
replace the sixteen screws (0515-0372). Torque to 9 inch-pounds.
5. Replace the front panel. Refer to the Front Frame Assembly replacement
procedure.
6. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.

Table of Contents

Other manuals for Keysight Technologies N9030B

Related product manuals