104 Keysight B2980B User’s Guide, Edition 1
How to Perform Measurement
Resistance Measurement
Pour fixer l'adaptateur N1413A/N1414A, enfoncez-le dans les connecteurs
B2985B/B2987B.
Poussez l'adaptateur fermement jusqu'à ce qu'il se verrouille en position (espace
< 1 mm).
For using the N1413A/N1414A adapter, set the Low Terminal State to FLOATING.
Voltage Source Low is connected to the chassis ground through the adapter.
If you use the N1413A/N1414A adapter, you do not need to care about setting of
the Voltmeter connector’s inner shield. It is opened at the adapter’s input.
Figure 3-12 Floating Device Measurement
Voltage Source Low terminal is internally connected to or disconnected from the
circuit common as shown in Figure 3-12. See “Low Terminal State” on page 110
for more information.