EasyManuals Logo

Keysight DAQ970A User Guide

Keysight DAQ970A
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
Secure - Enables (On) or disables (Off) the instrument security. When On, NISPOM Sanitize softkey will
appear and security must be unlocked to make changes to I/O enables and configuration.
NISPOMSanitize sanitizes all user-accessible instrument memory except for the calibration constants and
reboots the instrument. This complies with requirements in chapter 8 of the National Industrial Security
Program Operating Manual (NISPOM). When Off, I/O changes can be made without unlocking security.
The
NISPOM Sanitize
softkey and the SYSTem:SECurity:IMMEdiate command are equivalent. They
are for customers, such as military contractors, who must comply with NISPOM.
This feature destroys all user-defined state information, measurement data, and user-defined I/O set-
tings such as the IP address. This feature is not recommended for use in routine applications
because of the possibility of unintended data loss.
Admin
Press Firmware Update to update the instrument firmware to a new version. See Firmware Update details.
Keysight DAQ970A User's Guide 151
[Channel] Menu Overview

Table of Contents

Other manuals for Keysight DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ970A and is the answer not in the manual?

Keysight DAQ970A Specifications

General IconGeneral
Scan RateUp to 450 channels/sec
ConnectivityUSB, LAN, GPIB (optional)
Display4.3-inch color LCD
Operating Temperature0 °C to 50 °C
Number of slots3 slots
Measurement TypesDCV, ACV, DCI, ACI, 2-wire resistance, 4-wire resistance, frequency, temperature, capacitance
Voltmeter6.5 digit
Resolution22 bits