Press [View] on the front panel to view the scanned memory readings, alarm queue, error queue, and num-
ber of relay cycles during scanning or monitoring mode on the selected channel.
Views scanned memory readings (List, Trend Chart, Histogram, Statistics)
In View menu page, press Display to select the display mode as shown below:
List
Trend Chart
Histogram
Statistics
List
Displays the scanned memory readings in list format. Press Browse to display the readings starting from
the beginning of scan (First Page), or starting from the most recent sweep (Last Sweep). Use the knob or
the front panel arrow keys to view the list on subsequent pages; use the right/left arrow keys to move or
return the readings to the last/first page.
Trend Chart
3 Features and Functions
82 Keysight DAQ970A User's Guide

Table of Contents

Other manuals for Keysight DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ970A and is the answer not in the manual?

Keysight DAQ970A Specifications

General IconGeneral
Scan RateUp to 450 channels/sec
ConnectivityUSB, LAN, GPIB (optional)
Display4.3-inch color LCD
Operating Temperature0 °C to 50 °C
Number of slots3 slots
Measurement TypesDCV, ACV, DCI, ACI, 2-wire resistance, 4-wire resistance, frequency, temperature, capacitance
Voltmeter6.5 digit
Resolution22 bits
MemoryNon-volatile memory