Gocator Line Profile Sensors: User Manual
Gocator Web Interface • 159
You can turn regions off by unchecking the checkbox next to the Regions setting. When you do
this, the measurement tool uses the entire active area.
All tools provide region settings under the Parameters tab.
Region settings are often found within expandable feature sections in the tool's panel.
The Region settings apply to all of a tool's measurements.
See the individual tools for any details on using this parameter with each tool.
This parameter is also referred to as a measurement region.
In 2D mode, the tool region defaults to the center of the current data view, not the global field
of view. In 3D mode, the region defaults to the global field of view.
To configure regions:
1. Go to the Measure page by clicking on the Measure icon.
The scan mode must be set to the type of measurement you need to configure.
Otherwise, the wrong tools, or no tools, will be listed on the Measure page.
2. In the Tools panel, click on a tool in the tool list.
3. Expand the region section by clicking on the expand button .
Some region settings are found within other settings in this area.
4. Configure the region using the fields or graphically using the mouse in the data viewer.
Decisions
Results from a measurement can be compared against minimum and maximum thresholds to generate
pass / fail decisions. The decision state is pass if a measurement value is between the minimum and
maximum threshold. In the data viewer and next to the measurement, these values are displayed in
green. Otherwise, the decision state is fail. In the user interface, these values are displayed in red.
All measurements provide decision settings under the Output tab.