Gocator Line Profile Sensors: User Manual
Getting Started 33
3. Attach each of the two DINrail mount clips to the back of the Master using an M4x8 flat socket cap screw for
each one.
The following illustration shows the installation of clips on a Master 810 for horizontal mounting:
Ensure that there is enough clearance around the Master for cabling.
Grounding: Master Network Controllers
The rack mount brackets provided with all Masters are designed to provide adequate grounding through
the use of star washers. Always check grounding with a multi-meter by ensuring electrical continuity
between the mounting frame and RJ45 connectors on the front.
When using the rack mount brackets, you must connect the frame or electrical cabinet to which
the Master is mounted to earth ground.
You must check electrical continuity between the mounting frame and RJ45 connectors on the
front using a multi-meter.
If you are mounting Master 810 or 2410 using the provided DIN rail mount adapters, you must ground
the Master directly; for more information, see Grounding When Using a DIN Rail (Master 810/2410)
below.
Grounding When Using a DIN Rail (Master 810/2410)
If you are using DIN rail adapters instead of the rack mount brackets, you must ensure that the Master is
properly grounded by connecting a ground cable to one of the holes indicated below. The holes accept
M4x5 screws.