Gocator Line Profile Sensors: User Manual
Troubleshooting 464
l
Check that the exposure time is set to a reasonable level.See Exposure on page 98 for more information on
configuring exposure time.
Performance
The sensor CPU level is near 100%.
l
Consider reducing the speed. If you are using a time or encoder trigger source, see Triggers on page 88 for
information on reducing the speed. If you are using an external input or software trigger, consider reducing
the rate at which you apply triggers.
l
Consider reducing the laser profile resolution.
See Spacing on page 102 for more information on configuring resolution.
l
Review the measurements that you have programmed and eliminate any unnecessary measurements.