Gocator Line Profile Sensors: User Manual
Specifications 510
Specification Value
Power Draw (Min.) 5.76 W
Safety Input Voltage Range +24 VDCto +48 VDC
Encoder Signal Voltage Differential (12 VDC)
Digital Input Voltage Range Logical LOW: 0 to +0.1 VDC
Logical HIGH: +3.3 to +24 VDC
When using a Master hub, the chassis must be well grounded.
The power supply must be isolated from AC ground. This means that AC ground and DC
ground are not connected.
The Power Draw specification is based on a Master with no sensors attached. Every sensor has
its own power requirements that need to be considered when calculating total system power
requirements.