Gocator Line Profile Sensors: User Manual
Specifications • 513
Function Pin
Power In+ 1
Power In+ 2
Power In- 3
Power In- 4
Safety Control+ 5
Safety Control– 6
Power and Safety (6 pin connector)
The +24-48VDCpower supply must be isolated from AC ground. This means that AC ground
and DC ground are not connected.
The Safety Control requires a voltage differential of 24 VDC to 48 VDC across the pin to enable
the laser.
Function Pin
Input 0 Pin 1 1
Input 0 Pin 2 2
Reserved 3
Reserved 4
Reserved 5
Reserved 6
Reserved 7
Reserved 8
GND (output for powering other devices) 9
+5VDC (output for powering other devices) 10
Input (10 pin connector)
The Input connector does not need to be wired up for proper operation.
For Input connection wiring options, see Input on page 516.
Function Pin
Encoder_A_Pin_1 1
Encoder_A_Pin_2 2
Encoder_A_Pin_3 3
Encoder_B_Pin_1 4
Encoder_B_Pin_2 5
Encoder (11 pin connector)