12-6 DICOM
DICOM storage setting items are described as follows:
Name Description
Device
After you set the server(s) in DICOM Server Setting, the name(s) will
appear in the drop-down list, select the name of the storage server.
Service Name Default is xxx-Storage, and it can be modified.
AE Title
Application Entity title, here, it should be consistent with that of the
storage server.
Port
DICOM communication port, 104 by default. Here, the port should be
consistent with that of the storage server port.
Maximum
Retries
Refers to times that the system performs retrying connection, the
adjusting range is 0-9.
Interval Time(s)
Reserved feature.
Timeout
Refers to time after which the system will stop trying to establish a
connection to the service. Value: 560s.
Allow
Multiframe
If SCP supports this function, then select it.
Max Framerate Set the frame range of transferring cin file into DCM multi-frame file.
SR Storage
Option
To enable or disable structured reporting sending.
Cine Zoom
Mode
Select the cine zoom mode during image file storage.
Compression
Mode
Select the compression mode: uncompressed, RLE, JPEG and
JPEG2000.