Acoustic Output 17-5 
17.7  Acoustic Output 
17.7.1  Derated Ultrasonic Output Parameters 
In order to determine the relevant Ultrasonic Output Parameters, a method is used which 
allows for the comparison of ultrasound systems which operate at different frequencies 
and are focused at different depths. This approach, called "derating" or "attenuating", 
adjusts the acoustic output as measured in a water tank to account for the effect of 
ultrasound propagation through tissue. By convention, a specific average intensity 
attenuation value is used, which corresponds to a loss of 0.3 dB/cm/MHz. That is, the 
intensity of ultrasound will be reduced by 0.3 dB/MHz for every centimeter of travel from 
the transducer. This can be expressed by the following equation: 
)10/3.0-(
10
zf
wateratten
c
II
××
×=
 
Where I
atten
 is the attenuated intensity, I
water 
is the intensity measured in a water tank (at 
distance z), fc is the center frequency of the ultrasound wave (as measured in water), and 
z is the distance from the transducer. The equation for attenuating pressure values is 
similar except that the attenuation coefficient is 0.15 dB/cm/MHz, or one-half the intensity 
coefficient. The intensity coefficient is double the pressure coefficient because intensity is 
proportional to the square of pressure. 
Although the attenuation coefficient chosen, 0.3 dB/cm/MHz, is significantly lower than 
any specific solid tissue in the body, this value was chosen to account for fetal 
examinations. In early trimester ultrasound fetal examinations, there may be a significant 
fluid path between the transducer and the fetus, and the attenuation of fluid is very small. 
Therefore the attenuation coefficient was lowered to account for this case. 
17.7.2  Limits of Acoustic Output 
In accordance with the FDA Track 3 requirements, the derating (or attenuated) approach 
was incorporated into the FDA Acoustic Output Limits, as listed below. The maximum 
acoustic output level from any transducer in any operating mode is expected to fall below 
these limits. 
FDA Maximum Acoustic Output Limits for Track 3 (Attenuated Values) 
Application 
I
spta.3
 
(mW/cm
2
) 
I
sppa.3
 
(W/cm
2
) 
or 
MI 
Regions   
(except eyes)
720  ≤ 190  ≤ 1.9 
17.7.3  Differences between Actual and Displayed MI 
and TI 
In operation, the system will display to the operator the Acoustic Output Parameters 
Thermal Index, TI, or Mechanical Index, MI (or sometimes both parameters 
simultaneously). These parameters were developed as general indicators of risk from 
either thermal or mechanical action of the ultrasound wave. They serve to indicate to the 
operator whether a particular setting of the system increases or decreases the possibility