System Preparation 3-3
 Turning on the Power
To turn on the system:
1. Press the power button in the upper right corner on the control panel
2. The system enters the work status.
3. The indicator lights on and becomes green.
 To check the system after it is turned on
Press the power button in the upper right corner on the control panel to turn on the
system.
To check the system after the system is turned on:
No.
Check Item
<1> There shall be no unusual sounds or smells indicating possible overheating.
<2> There shall be no persistently displayed system error message.
<3>
There shall be no evident excessive noise, discontinuous, absent or black artifacts
in the B mode image.
<4>
Check if there is abnormal heat on the surface of the probe during an ultrasound
procedure.
<5> The control panel keys and knobs are fully functional.
<6>
The exam date and time are the same as the system date and time, and are
displayed correctly.
WARNING:
1.
If you use a probe giving off excessive heat, it may burn
the patient.
2.
If you find anything not functioning properly, this may
indicate that the system is defective. In this case, shut
down the system immediately and contact Mindray
Customer Service Department or sales representative.
NOTE: When you start the system or switch between probes, you will hear clicking
sounds – this is expected behavior.
3.3.2 Powering OFF the System
You need to follow the correct procedures to power off the system. In addition, after you
upgrade the software or when the system is down, you need to power off and restart it.
If you will not use the system for a long period of time, you shall:
1. Disconnect the power adapter.
2. Disconnect the mains power.
3. Turn off powers of all peripherals connected to the system.
 To power off your system normally:
Gently press the power button once on the upper right corner of the control panel. The
[Shutdown Confirm] screen appears. As the following: