3-4 System Preparation 
 
 
Shut down: To power off the system normally. 
 
Standby Click to enter the standby status. 
 
Cancel: to cancel the operation. 
 
  To shut down the system in a direct way if you cannot do it normally: 
Press and hold the power button for a long time and the system will power off without 
displaying the [Shutdown Confirm] screen. However, shutting down the system this way 
may destroy the data. 
NOTE:
1.
DO NOT rush shutdown of the system in a direct way. It may destroy the 
data. 
  2.
After the software upgrade, please shut down the system in the normal way 
(using “Shut down” method) so as to guarantee a fully update. 
3.3.3  Standby 
Definition of standby: in a normal environmental temperature and humidity condition, the 
system is connected with only one probe while no other external devices or extend 
modules are connected, only 3V3_STB and 5V_STB power supply is available and in the 
meantime, the standby indicator is in orange. 
When the battery capacity is charged to the full capacity, the standby time of the system is 
no less than 64 hours. 
  To enter standby: 
 
Fully fold the LCD display and wait for 30 sec, then the system enters the standby 
status. 
 
Open [Setup]→[System Preset]→“General” to set the time for screensaver and 
standby. If there is no operation, the system enters screensaver after the waiting 
time set for screensaver, and if there is still no operation, the system enters 
standby after the waiting time for standby. 
 
Press the power button and select “Standby”. 
When the system enters the standby status, the power button backlight turns orange. 
  To exit standby 
 
Unfold the LCD display. 
 
Press the power button. 
  When the system enters the standby status, if need to power off: 
 
Long press the power button for 8s. 
 