Page #259 background image
Probes and Biopsy 13-13
Refer to the instructions provided by the chemical manufacturer concerning
concentration of the sterilization solution, method of sterilization and dilution and
cautions during use.
Do not soak the probe connector or the cable near it into water or any solution.
Follow local regulations when selecting and using the sterilization solution.
3. Rinse the probe with plenty of sterile water (about 2 gallons) for at least 1 minute to
remove all chemical residues on it. Or, follow the rinsing method recommended by the
sterilization solution manufacturer to rinse the probe.
4. Wipe off the water on the probe with sterile cloth or gauze after rinsing it. Do not dry
the probe by heating.
Immerse the probe in the solution (take 7LT4s as an illustration)
NOTE:
Repeated disinfection may degrade the performance and safety of the
probe.
13.1.6 Storage and Transportation
When all examinations for the day have been completed, confirm that the probe is in good
condition. After disinfecting the probe, confirm that the probe is in good condition and
store it in a suitable place so that the next examination can be conducted smoothly.
1. To prevent the probe from being damaged, DO NOT store it in locations where it may
be exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
2. Store and transport the probe under the following ambient conditions:
Ambient temperature: -20°C 55°C
Relative humidity: 30% 95% (no condensation)
Atmospheric pressure: 700 hPa 1060 hPa
Store and transport the 4CD4s probe under the following ambient conditions:
Ambient temperature: -10°C 60°C
Relative humidity: 30% 95% (no condensation)
Atmospheric pressure:700 hPa 1060 hPa
3. When the probe is sent to MINDRAY Customer Service Department or sales
representative for repair, be sure to disinfect it and keep it in the carrying case to
prevent infection.
4. Sterilize the carrying case as necessary.