DICOM 12-15 
 
DICOM query/retrieve setting items are described as follows: 
Device 
After you set the server (s) in DICOM Server Preset screen, the name (s) will 
appear in the drop-down list, select the name of the server. 
Service 
Name 
Default: xxx-Store, and it can be modified 
AE Title  Application Entity title 
Port 
DICOM communication port, 104 is default. Here, the port should be 
consistent with that of the query/retrieve server port. 
Maximum 
Retries 
Reserved feature. 
Interval 
Time(s) 
Reserved feature. 
Timeout 
Value: 5-60s, in increments of 5s, and 15 by default. 
Add  Add the DICOM service to the service list 
Cancel  Click to cancel parameter setting 
Delete  Click to delete the selected service in the service list 
Update 
Select an item in the service list, change the parameters in the above area, 
and click [Update] to update the item in the service list. 
Verify  Click to verify if the two DICOM application entities are normally connected. 
Default 
Select an item in the service list, click [Default] and you can see “Y” in the 
Default column. 
Exit  Click to exit the screen.