System Overview 2-11 
UMT-300 
WARNING:
1.
Be sure to connect the equipotential wire before 
inserting the power plug into the receptacle; be sure to 
remove the power plug from the receptacle before 
disconnecting the equipotential wire; otherwise electric 
shock may result. 
 
2.
When you connect another device to the trolley, you 
should use the equipotential wire to connect each of 
equipotential terminals; otherwise electric shock may 
result. 
 
3.
Connect the earth cable before powering ON. Disconnect 
the earth cable after powering OFF. Otherwise, electric 
shock may result. 
 
4.
DO NOT connect the trolley to the outlets with the same 
circuit breakers and fuses that control the current to 
devices such as life-support systems. If the trolley or the 
system malfunctions and generates over-current, or 
when there is an instantaneous current at power ON, the 
circuit breakers and fuses of the building’s supply circuit 
may be tripped. 
 
5.
If UMT-300 has configured with display module, please 
take care of the display module position to leave enough 
space below when adjusting the display support arm, so 
that the display module will not hit into the monitor of the 
ultrasound system. 
 
CAUTION:
1.
When moving the trolley with mounted system, please 
take care of the connector of the power adapter in case of 
damage. 
  2.
Maximum output power of the outlet in the trolley is 
350VA. 
 
NOTE
1.
If UMT-300 has configured with display module, please take care of the 
display module position in case of hurting one’s head. 
2.
If probe extend module is configured, please do not place any objects under 
the module.