EasyManua.ls Logo

MR ETOS IM - Page 214

MR ETOS IM
412 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Operation
Maschinenfabrik Reinhausen GmbH 2019214 5163667/06 ENETOS
®
IM
Figure161: Configuring bushing monitoring
F1/F2-C: Activate capacitance monitoring
You can use this parameter to activate or deactivate capacitance monitoring
in field 1 or field 2.
After being switched on, capacitance monitoring only becomes active once
a delay time has elapsed (factory setting of 2minutes) in order to avoid er-
rors resulting from transients.
To activate/deactivate capacitance monitoring, proceed as follows:
1. Go to Settings > Parameters > Bushing monitoring field1/field2 > F1/
F2-C: Activate capacitance monitoring.
2. Select the desired option.
3. Press the Accept button to save the modified parameter.
F1/F2-C: C1 Phase L1/L2/L3
You can use these parameters to set the reference capacitance C1 for the
bushings of phases L1, L2 and L3 in field 1 and field 2. The reference value
is the value you have measured with an external measuring device for com-
missioning.
To set the reference capacitance, proceed as follows:
1. Go to Settings > Parameters > Bushing monitoring field1/field2 > C:
C1 phase L1/L2/L3.
2. Enter the desired value.
3. Press the Accept button to save the modified parameter.

Table of Contents