12 Technical data
Maschinenfabrik Reinhausen GmbH 2019400 5163667/06 ENETOS
®
IM
12.12 Standards and directives
Electrical safety
IEC 61010-1
IEC 61010-2-030
Safety requirements for electrical measurement and con-
trol and regulation equipment and laboratory instruments
▪ Protection class 1
▪ Overvoltage category III
▪ Contamination level 2
IEC 60950-1 Information technology equipment – Safety
Table159: Electrical safety
Electromagnetic compatibility
IEC 61000-4-2 Immunity to electrostatic discharge (ESD)
▪ Front panel and operating elements
▪ Contact: ±8 kV
▪ Air: ±15 kV
▪ Terminals, plug connectors, and interfaces:
▪ Contact: ±6 kV
▪ Air: ±8 kV
IEC 61000-4-3 Immunity to high-frequency electromagnetic fields
▪ 20V/m; 80...4000MHz; 80% AM
▪ 20V/m; 900MHz±5MHz; PM
IEC 61000-4-4 Immunity to quick, transient electrical disturbances (burst)
▪ Power supply: 4 kV
▪ Measurement (UI1/3): 4 kV
▪ Digital I/O: 4 kV
▪ Analog I/O, shielding on both sides: 4 kV
▪ Communication interfaces, shielding on both sides: 4
kV
IEC 61000-4-5 Immunity to impulse voltages (surge)
▪ AC power supply: 4 kV CM, 2 kV DM
▪ DC power supply: 2 kV CM, 1 kV DM
▪ Measurement (UI1/3): 4 kV CM, 2 kV DM
▪ Digital I/O: 2 kV CM, 1 kV DM
▪ Analog I/O, shielding on both sides: 2 kV CM
▪ Communication interfaces, shielding on both sides: 2
kV CM
IEC 61000-4-6 Immunity to conducted disturbances, induced by high-fre-
quency fields
▪ 10 V, 150 kHz...80 MHz, 80% AM