EasyManua.ls Logo

ORTEC radEAGLE - Page 132

ORTEC radEAGLE
147 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D.4 Checking the Proper Function of the System radEAGLE User Manual
D.4 Checking the Proper Function of the System
To ensure your radEAGLE is working properly, we will supply a short checklist for successful
operation.
1. Check the status of the battery failure LED
a) After some time, the booting screen of the radEAGLE should appear.
If the screen does not appear, check if the display has backlight. If not, there
might be a problem with the battery. Power the instrument with external
power and check whether the problem persists.
If the instrument boots with power cable connected, check the status of
the battery switch in the rear battery chamber of the instrument.
b) Is the orange battery fault LED on?
If yes, there might be charging problem or some other problem with the
batteries. If it is running, turn off the instrument, and try charging the bat-
teries.
2. System boot-up and welcome screen
a) Self-checking routines run in the background of the boot process. If a self-check
fails, a corresponding error message will appear on the device and give you fur-
ther advice.
b) Once started, the system should welcome you in dose rate mode. If no source
is around, the ambient dose rate is expected to be between 0.01 µSv/h and
0.08 µSv/h.
c) The count rate in cps should be greater than 0. There are always natural radia-
tion counts.
d) If you have a neutron detector, the neutron cps should be close to 0.00 cps.
Sometimes values of about 0.05 might occur. If you observe a neutron count
rate of 0.5 cps or greater, it is likely that a neutron source is nearby.
e) If the battery was charged, the battery status bar should indicate fully charged
status
f) If the bar shows a low battery, this might point towards a problem with the bat-
teries. Try charging the battery again.
3. Specific checks in spectrum mode
a) Enter spectrum mode. Without a source, start a spectrum acquisition and ob-
serve the area around 1460 keV. After a few minutes, the natural potassium
peak should appear at 1460 keV. You can use this peak to verify the correct po-
sitioning of the instrument even if no cesium calibration source is available. After
fresh startup, the instrument should have at least a precision of around ±0.5 %
of the line energy or a maximum deviation of ±7 keV around the 1460 keV line.
132/147 Software 3.2.12 Document 3.3.0o © innoRIID GmbH 2019-04-11

Table of Contents