5.14 Easy Calibration radEAGLE User Manual
After you let the radEAGLE know which calibration source is in front of it (Fig. 67, p. 64), it
will aquire data for a while and use it for calibration (Fig. 68, p. 64).
The acquired spectrum is shown, overlayed by the theoretical peak position of the selected
source and the current peak fit of the measured spectrum.
Figure 67: Selecting a source for easy calibration
Figure 68: Easy calibration in progress
A progress percentage is shown during the data collection (Fig. 69, p. 65), sometimes in sev-
eral cycles, until the peak position of the calibration source is within ±0.5 % of the theoretical
position.
Settings and Commands
Exit
Return to the superior screen.
Down
Highlight the next item. Cycles to the top item after you reached the last item.
Select
Select the highlighted item.
64/147 Software 3.2.12 • Document 3.3.0o © innoRIID GmbH • 2019-04-11