LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Printer Control Keypad (A09) Replacement
Section Menu Section Contents Procedures Back Index
199
8
Removing the Printer Control Keypad (A09)
Refer to Figure 8.15 on p. 198.
To remove the printer control keypad from outside the front case:
1. Using a very thin, wide, flat-edged tool (example, 1.5” putty knife), gently pry one edge of the printer control keypad until it is
released from its adhesive mount. Take care not to damage the case.
2. Pull the printer control keypad away from the case, extending part of the W12 cable (3206989-000) (see Figure 9.52 on p. 474)
through the keypad opening.
3. Disconnect the W12 cable from printer control keypad at J41. Make sure the W12 cable does not fall back into the front case.
Installing the Printer Control Keypad (A09)
Refer to Figure 8.15 on p. 198.
NOTE: Before installing the new printer control keypad, verify that the shelf-life date printed on the printer control keypad package
has not expired.
To install the new printer control keypad (see Table 9.11 on page 418 for keypad part numbers):
1. Using a soft, lint-free cloth and isopropyl alcohol, gently remove old adhesive from the printer control keypad cavity on the front
case.
2. Connect the W12 cable (3206989-000) to the printer control keypad at J41.
3. Check that connector locking tab is securely latched.
4. Remove the protective covering from the keypad adhesive surface and press the keypad firmly and evenly into the keypad cavity
on the front case.