2. From the outside of the front case, insert the new therapy Connector cable
through the therapy mounting seal and front case.
3. Secure the therapy connector to the front case with four new screws (3207361-
375); torque to 10 in-lb using a T15 bit.
4. Reassemble the case as described in Reassembling the Case (p. 184).
To install the therapy Connector cable into the front case:
(Continued) 4 steps, (Page 2 of 2)