Replacement Procedures
System (A01)/Therapy (A04) PCB Assembly Replacement
Section Menu Section Contents Procedures Back Index
238
8
8. Lift the system/therapy board assembly out from the rear case.
Turn the system/therapy board assembly over to expose the
therapy board.
9. Immediately discharge the pacing capacitor as described in
Discharging the C15 Pacing Capacitor (p. 179). The discharge
points are located at resistors R5 and R20 on the therapy board.
To remove the system/therapy PCBs as a single unit from the rear case: (Continued) 9 steps, (Page 6 of 6)