LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Installing the System (A01)/Therapy (A04) PCB Assembly
Section Menu Section Contents Procedures Back Index
252
8
8. Follow the instructions to install the system/therapy assembly into device as described in Installing the System (A01)/Therapy
(A04) PCB Assembly (p. 239).
9. After device reassembly, complete the TCP – Defibrillator Energy Calibration procedure. The defibrillator calibration constants
may be invalidated when you replace Therapy PCB.
To replace the therapy PCB (A04): (Continued) 9 steps, (Page 3 of 3)