LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Printer Assembly (A12) Replacement
Section Menu Section Contents Back Index
361
8
Installing the Printer Assembly (A12)
NOTE: The new printer comes with new screws installed to aid reassembly.
To install the printer assembly into the front case.
1. Lay the device face up on the battery wells.
2. Insert the printer into the front case, and then gently push toward the therapy connector (to engage the printer connector)
3. Tighten the two screws from inside the printer assembly; torque to 10 in.lb using a P2 bit.
4. Tighten the third screw on the outside left side of printer; torque to 10 in.lb.
5. Apply a new screw cover label. Apply the screw cover label (#6) from the label set (3207318-XXX) to the left front of printer.
6. Install printer paper.