Chapter F - ADJUSTMENTS
Planmeca Compact i F-37
WMS OPERATION
Technical Manual
Cleaning agent is fed to internal water lines
City Water Bottled Water
Feeding cycle:
Internal water lines
Similar to City water mode
V1 c
V2 o
V3 o
V4 o
V5 c
Pump running
Display countdown timer 30 s
Duration 30 s
Excess cleaning agent is removed from bottle
City Water Bottled Water
Feeding cycle:
Emptying bottle
Similar to City water mode
V1 c
V2 c
V3 o
V4 o
V5 c
Pump not running
Display wait, when L2 reached countdown
timer 15 s
Duration until L2 reached + 15 s
When feeding cycle is completed, Display: He 66
Flushing cycle is started
City Water Bottled Water
After 8 hours if unit let on or when unit restarted
on
When unit turned on

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact I and is the answer not in the manual?

Planmeca Compact I Specifications

General IconGeneral
SpittoonYes
Operating LightLED
Imaging TechnologyDigital
X-ray Tube Voltage60-70 kV
X-ray Tube Current7 mA
Focal Spot Size0.8 mm
Exposure Time Range3.2 s
Power Supply50/60 Hz