Chapter C - SERVICE MODE
C-46 Planmeca Compact i
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n 207 Headrest length in millimetres (mm) in relation to
calibration point and virtual origo point
This service mode shows the headrest length in millimetres (mm) in relation to calibration
point and virtual origo point.
n 208 Headrest height in millimetres (mm) in relation to backrest
zero-level and virtual origo point
This service mode shows the headrest height in millimetres (mm) in relation to backrest zero-
level and virtual origo point.
n 209 Headrest angle in degrees (deg) in relation to backrest
angle and virtual origo point
This service mode shows the headrest angle in degrees (deg) in relation to backrest angle
and virtual origo point.
n 300 Headrest communication status
1. (leftmost) digit: Always 1
2. digit: Output message
3. digit: Drive state
4. (rightmost) digit: Input mode
Output messages:
0: Power is off
1: Power is on, slowing down (no control)
2: Power is on, responding to buttons
3: Power is on, automatic drive
Drive states:
0: Headrest not installed
1: Headrest stopped (emergency)
2: Headrest ready
3: Headrest slowing down
4: Headrest running, manual operation
5: Headrest running, automatic operation
Input modes:
0: Headrest input unknown
1: Headrest not installed or open circuit
2: Headrest short circuit
3: Headrest input wrong CRC
4: Headrest input unknown mode
5: Headrest input normal mode but unknown type
6: Headrest input in burn mode but unknown type
7: Headrest input in burn mode
8: Headrest input normal
1028 is typical idle value.
1248 is typical manual drive value.