Chapter C - SERVICE MODE
Planmeca Compact i C-45
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n. 137 Bowl rotation safety features
In this mode the bowl rotation safety operation (when chair is driven upwards) is defined. If
the dental unit is not equipped with bowl, the bowl rotation is not included in the safety circuit.
0 = Bowl rotation is not included in the safety circuit
1 = Bowl rotation is included in the safety circuit (default)
n.138 Simulation system type
Use this mode to select the Simulation system type.
0 = IDSS - Interactive Digital Simulation System
1 = ASW-1 - Advanced Simulation Workplace Single (Main Valves are in normal use)
2 = ASW-N - Advanced Simulation Workplace (Main Valves are not in use)
n.139 Duration of long instrument hose flushing
Use this mode to set the duration of the long instrument hose flushing. The adjustment range
is 2 - 9 minutes. Factory default is 2 minutes.
n.140 Suction valve closing time adjustment
Use this mode to set the suction valve closing time. The adjustment range is 0 - 3 seconds.
Factory default is 0.
n 200 Boot information bits
Reserved for factory use.
n 201 Boot software checksum
Reserved for factory use.
n 202 Main software checksum
Reserved for factory use.
n.203 Processor watchdog test
Restarts the dental unit with error E 75.1.
n.204 Processor reset test
Restarts the dental unit without errors.
n.205 Lifter height calibration in millimetres (mm)
Measure from floor to the center of the backrest joint and set that value to this service mode.
n.206 Backrest angle calibration in degrees (deg)
Measure angle (or drive to 0 with spirit level) and set that angle here.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact I and is the answer not in the manual?

Planmeca Compact I Specifications

General IconGeneral
SpittoonYes
Operating LightLED
Imaging TechnologyDigital
X-ray Tube Voltage60-70 kV
X-ray Tube Current7 mA
Focal Spot Size0.8 mm
Exposure Time Range3.2 s
Power Supply50/60 Hz

Related product manuals