Chapter C - SERVICE MODE
C-48 Planmeca Compact i
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n.312 Headrest length ADC-value
Value is typically from 178 to 954.
After pressing the S key, the length can be driven.
n.313 Headrest joint 2 (near cushion) drive WITHOUT limits and
potentiometer in Set mode
In this mode the headrest joint 2 (near cushion) is driven WITHOUT limits and potentiometer
in Set mode.
Displays text “H2- ” on the display in Set mode.
n.314 Headrest joint 1 (near backrest) drive WITHOUT limits and
potentiometer in Set mode
In this mode the headrest joint 1 (near backrest) is driven WITHOUT limits and potentiometer
in Set mode.
Displays text “H1- ” on the display in Set mode.
n.315 Headrest length calibration
Like service modes SM n.79 and SM n.80. Shows text “HL--“. When potiontiometer value is
too far, rightmost block is up, when potentionmeter value too near, rightmost block is down.
When potentiometer is calibrated, the rightmost block is on the middle. Headrest length joint
has to be driven first on the outer limit, then tune the potentiometer according the rightmost
blocks.
After pressing the S key, the length can be driven WITHOUT limits and potentiometer.
n 316 Headrest joystick directions and buttons
1. block (leftmost) spare I/O
2. block left joystick button
3. block right joystick button
4. block toggle from any button
5. block joystick out
6. block joystick in
7. block joystick down
8. block (rightmost) joystick up
n 317 Potentiometer select line test.
If the value is 0 (or +-1) and will not change when the joystick(s) are moved, the select wire is
broken.
n.318 As service mode SM n.18 but only headrest endurance
test
Reserved for factory use.
n.319 As service mode SM n.18 but chair and headrest
endurance tests.