3. Installation and setup ABL800 FLEX Operator's Manual
Quality control setup, Continued
QC ranges
(continued)
Step Action
4.
Highlight the limit by touching it on the screen and enter your own
value(s), using the keypad.
5.
To activate or deactivate the Fixed SD (i.e. a minimum allowed
control range is defined by setting a Fixed SD), press the check
button.
To change the SD value, touch the SD field to highlight it and enter
the value, using the keypad. Confirm the entry with Enter.
6.
Repeat the procedure for other parameters in the same manner.
7.
To update the control ranges for all parameters of a displayed level to
the corresponding lot-to-date (2 SD) range shown, display the desired
slot, using Next Slot.
Press Update All.
8.
Press Continue to update the control ranges for all parameters under
the specified solution slot.
Press Cancel to cancel updating.