ABL800 FLEX Operator's Manual 3. Installation and setup
Quality control setup, Continued
NOTICE: Once the Fixed SD is activated, you cannot update the control ranges to
the limits that are narrower than those determined by the Fixed SD, for both single-
parameter and multiple-parameter updates.
QC ranges
(continued)
This program allows you to select the following for the Quality Control
Identification screen during a measurement:
QC input setup
mandatory temperature entry by the operator
the default temperature always displayed (unless altered by the operator).
To select the settings, do the following:
Step Action
1.
Activate the Mandatory temperature check button.
A
"
will appear next to an empty temperature box on the Quality
Control Identification screen during each quality control
measurement and a temperature must be entered, otherwise the result
cannot be retrieved.
2.
Highlight the default temperature in the "Default temperature" box
and enter a default temperature on the keypad; confirm with Enter.
The value in °C or °F is automatically entered on the Quality Control
Identification screen during measurement. The temperature can be
altered for a particular measurement but will return to the default
setting for future measurements.
Continued on next page
3-39

Table of Contents