3. Installation and setup ABL800 FLEX Operator's Manual
Quality control setup, Continued
Westgard Rules
(continued)
Step Action
2.
Press Next Slot and activate/deactivate the Westgard Rules for the
other slots in the same manner.
3.
To activate the Westgard rules for specific parameters in a slot,
display the desired slots by using Next Slot.
4.
Press the Edit button. The screen shows all the Westgard Rules and
their current settings for the parameter.
5.
Display the desired parameter using the Next Param or Prev Param
buttons.
Parameter
6.
Activate the desired Westgard Rule(s) by pressing the corresponding
check button.
Once edited, all future quality control measurements on the parameter
for the given slot will be evaluated according to the selected Westgard
Rule(s).
7.
Select Westgard Rules for other parameters or levels in the same
manner.
8.
To select/deselect all six Westgard Rules for every parameter of a
displayed control solution slot, select the desired slot by pressing the
Next Slot button.
Continued on next page
3-44

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals