ABL800 FLEX Operator's Manual 3. Installation and setup
Quality control setup, Continued
QC Statistics
(continued)
Step Action
2.
To set up the analyzer to send reminders to print or export the monthly
statistics, activate the appropriate check button (or both) for the
message(s) to appear in System Messages upon occurrence.
Press Close to exit the program.
3.
Westgard Rules
The Westgard Rules are a set of statistical rules that, when applied to the quality
control results, can increase the probability of detecting an error in the sampling
procedure or in the analyzer itself, or help detecting a shift or trend in your quality
control results by comparing current measurement values of a control solution with
previous values.
This program allows you to select Westgard Rules for all slots or for specific
parameters.
To activate or deactivate the assigned Westgard Rules, do the following.
Step Action
Press On/Off button to activate the assigned Westgard Rules for the
slot.
1.
Slot, solution,
type, lot
Press On/Off again to deactivate the assigned Westgard Rules.
Buttons are
grayed out
Continued on next page
3-43