3. Installation and setup ABL800 FLEX Operator's Manual
Quality control setup, Continued
QC schedule
(continued)
To remove a quality control measurement from the schedule do the following:
Step Action
1.
Highlight the desired quality control measurement and press Remove.
2.
Press the desired button to remove the quality control measurement
from the schedule: For this day, For all days or All entries for QC
slot (number).
Press Back to return to the QC Schedule screen.
3.
QC Statistics
This program allows you to select the following:
• the statistics factor
• the initiation of warning messages at the end of the statistical month.
To select the settings, do the following:
Step Action
1.
Statistics factor expands the control range to the statistics range (it is a
range within which quality control results must fall in order to be
included in the quality control statistics).
Key in the desired statistics factor (from 1.0 to 9.9) on the keypad and
confirm with Enter. The default value is 1.5.
Continued on next page
3-42