EasyManuals Logo

RADIOMETER ABL800 FLEX User Manual

RADIOMETER ABL800 FLEX
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
3. Installation and setup ABL800 FLEX Operator's Manual
Quality control setup, Continued
Rilibäk Ranges
(continued)
To activate or deactivate the Rilibäk Rules, do the following.
Step Action
Press the On/Off button to activate/deactiviate the assigned Rilibäk
rules.
1.
To add a new Rilibäk rule, do the following:
Step Action
1.
Press the Add button to display the screen below:
2.
Select the desired parameter from the parameter list shown in the right
side of the screen.
3.
Press the Left arrow button until the first "Lower Limit" box is
highlighted and enter the desired lower limit.
4.
Highlight the next box and select "<" or "<=".
5.
Highlight the first "Upper Limit" box and select "<" or "<=".
6.
Highlight the next "Upper Limit" box and enter the desired upper
limit.
Continued on next page