ABL800 FLEX Operator's Manual 4. Sample measurements
General information, Continued
Available modes
and parameters
(continued)
Analyzer Measuring modes Measured parameters
805 (cont.) Ampoule QC All available
810 BG
only
Syringe – S85 μL
Capillary – C55 μL
pH, pCO
2
, pO
2
Capillary – FLEXMODE See below
Syringe – Pleura pH* pH
Other fluids (any mode) Depends on the selected
mode
Expired air pCO
2
, pO
2
Ampoule QC All available
810/20/30
Syringe – S85 μL
Capillary – C55 μL
pH, pCO
2
, pO
2
and
810: ctHb, sO
2
820: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb
830: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb, ctBil,
FHbF
Capillary – C35 μL OXI
810: ctHb, sO
2
820: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb
830: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb, ctBil,
FHbF
Capillary – FLEXMODE See below
Syringe – Pleura pH* pH
Other fluids (any mode) Depends on the selected
mode
Expired air pCO
2
, pO
2
Ampoule QC All available
* See page 4-5