4. Sample measurements ABL800 FLEX Operator's Manual
General information, Continued
Available modes
and parameters
(continued)
Analyzer Measuring modes Measured parameters
815/25/35
Syringe – S195/95 μL
Capillary – C95 μL
pH, pCO
2
, pO
2
, cK
+
, cNa
+
,
cCa
2+
, cCl
–
, cGlu, cLac and
815: ctHb, sO
2
825: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb
835: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb, ctBil,
FHbF
Syringe – S85 μL pH+BG +
OXI
Capillary – C55 μL
pH+BG+OXI
pH, pCO
2
, pO
2
and
815: ctHb, sO
2
825: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb
835: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb, ctBil,
FHbF
Capillary – C35 μL MET
cGlu, cLac
Capillary – C35 μL OXI
815: ctHb, sO
2
825: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb
835: ctHb, sO
2
, FO
2
Hb,
FCOHb, FMetHb, ctBil,
FHbF
Capillary – FLEXMODE See below
Syringe – Pleura pH* pH
Other fluids (any mode) Depends on the selected
mode
Expired air pCO
2
, pO
2
Ampoule QC All available
* See page 4-5
Continued on next page
4-4