11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
134 LIS/HIS:
Connection
established
The connection was
successfully
established.
No action required. The
message is informative.
165 LIS/HIS:
High level
protocol
could not
generate
high level
packet
An error occurred
while formatting a
message.
- Check protocol
configurations.
- Contact Radiometer
service representative.
166 LIS/HIS:
General
communica-
tion error
An internal error
occurred in the
LIS/HIS
communication
module.
Contact Radiometer service
representative if the
problem persists.
167 LIS/HIS:
High level
protocol
received
packet in
wrong
format
An error occurred
while parsing
(interpreting) a
message.
- Check protocol
configurations.
- Contact Radiometer
service representative.
200 User msg: This is only a message.
An operator has
entered a note in the
log.
No action required.
201 Westgard
Rule (1.2s)
violation
Measured parameter
value is outside the
mean ± 2 SD range.
202 Westgard
Rule (1.3s)
violation
Measured parameter
value is outside the
mean ± 3 SD range.
203 Westgard
Rule (2.2s)
violation
Two consecutive
measurements are
outside the mean ± 2
SD range on the same
side of the mean. This
may indicate a shift.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements (including
elctrodes).
- Refer to the Quality
Control Systems Reference
Manual for detailed
evaluation procedure.
Continued on next page